久久精品国产亚洲av麻豆,真人实拍女处被破的视频,91天仙tv国产福利精品
<menuitem id="1b5v3"><ruby id="1b5v3"></ruby></menuitem>
<cite id="1b5v3"></cite><var id="1b5v3"><noframes id="1b5v3"><ins id="1b5v3"></ins>
<ins id="1b5v3"><span id="1b5v3"></span></ins>
<cite id="1b5v3"><noframes id="1b5v3"><var id="1b5v3"></var>
<ins id="1b5v3"><noframes id="1b5v3"><cite id="1b5v3"></cite>
<cite id="1b5v3"><span id="1b5v3"></span></cite>
<ins id="1b5v3"><span id="1b5v3"><cite id="1b5v3"></cite></span></ins>
<ins id="1b5v3"></ins>
<ins id="1b5v3"></ins>
<ins id="1b5v3"></ins><ins id="1b5v3"><noframes id="1b5v3"><ins id="1b5v3"></ins>
<var id="1b5v3"></var>
<cite id="1b5v3"><noframes id="1b5v3">
<cite id="1b5v3"><span id="1b5v3"></span></cite>

M310A

M310A.jpg

FEATURES:

Support battery, capacitor material research, half-cell, three-electrode, button battery, small-capacity battery and other research tests, 0~5V full range controllable.

Supports precision research and testing with 0.01% voltage and current accuracy.

Support high-rate fast charge research, 30C can achieve 0.01% capacity accuracy; 100C can reach 0.1% capacity accuracy.

Support custom pulse test with a minimum pulse width of 2mS.

Full coverage of the measurement records of the voltage, current, capacity, energy, efficiency, capacitance, internal resistance and other parameters required for the study.

The device comes with UPS and storage to effectively prevent data loss and damage caused by downtime, sudden power failure, network failure, etc.


M310A PERFORMANCE:

ITEM

INSTRUMENT PERFORMANCE

Range

Voltage

5V (DC to -5V)*1

Current

100uA、1mA、10mA、100mA

ACCURACY

(SET,READ)

Voltage

±0.01%RD±0.01%FS

Current

±0.01%RD±0.01%FS

No. of CHANNELs


8Chl

No. of STEPs


1024

TIME

Accuracy

≤±5uS/S (Equipment autonomous control, no communication and other time delay errors)

Min Step Time

20mS

Min Pulse Time

2mS

Sampling Interval

≤10mS*2

INPUT IMPEDENCE


10GΩ(INPUT BASE CURRENT < 0.5nA

TEMPERATURE DRIFT


≤20ppm/

RESPONSE TIME


Current 0~90% ≤1mS; Current -90~90% ≤3mS

DATA RECORD

Sampling Rate

100HzSampling interval 10mS*2

Setting Method

Interval time, Sampling rate, Interval current, Interval capacity, Interval voltage, Interval energy

PROCESS CONTROL

Work Mode

REST/ CC / CV / CCCV / CRATE / CP / CR / CPULSE / CSLOPE / DC / DV / DCDV / DRATE / DP / DR / DPULSE / DSLOPE

End Conditions

TIME / VOLTAGE / CURRENT / POWER / CAPACITY / ENERGY / LOOPS / PHASE CYCLE

Protective Function

Power-off data protection, Net break protection

User-defined safety: Over voltage, Under voltage, Over current, Over rate, Over capacity.

FUNTION


DCIR、Capacitance capacity calculation、Step pluse、Three-electrode testReference test)、DQ/DV

OUTPUT


Four electrodes (support reference electrode test)

COMMUNICATION


10/100M Ethernet(RJ45)、USB(Console)

OPERATING ENVIRONMENT

Temperature

0~40℃(23±2Optimum accuracy

Humidity

90% No condensation

POWER SUPPLY


AC220V~±10% 50Hz / AC110V~±10% 60Hz,  40W *3

EQUIPMENT SIZE


1U  W*H*L:430*300*43.5(mm)

Note:    1. The voltage range can be customized by 2~8V.

2. The sampling rate can be customized to 1000Hz (sampling interval 1mS).

3. For best test accuracy and safety, be sure to ground well. The machine needs to be warmed for 30 minutes before use.

prev articleno more

next articleG310A

CLIENT

SERVICE

Consulting

South China East China North China Central China Overseas Other

Business Mailbox

tech@whland.com

Share to:

Top
国产午夜精品一区二区三区不卡,里番acg★里番本子全彩,精品一区二区三区无码免费视频